Tempe™

Tempe™ providing inaccurate temperature readings
RTD pins may be faulty In order to determine if Tempe™ RTD pins are faulty, perform the following procedure: Measure the resistance between pin A and ...
Thu, 29 May, 2014 at 4:17 PM
Tempe™ heater unresponsive
Heater pins may be faulty In order to determine if Tempe™ heater pins are faulty, perform the following procedure: Measure the resistance between pin ...
Fri, 30 May, 2014 at 10:18 AM
Crystals breaking prematurely
Problem Crystals breaking prematurely Cause Crystal not seated properly Solution Make sure that the crystal is seated properly inside the crystal co...
Fri, 30 May, 2014 at 11:04 AM
Weak Crystal Reading
Problem Crystal produces weak reading Cause Crystal contact spring may have become deformed Solution If the contact spring has become deformed, it ...
Fri, 30 May, 2014 at 11:11 AM
Insufficient Heat Conduction
Problem Heat conduction is insufficient to heat sensor head Cause The heater’s ceramic insulators may have become dislodged - which often occurs if the...
Fri, 30 May, 2014 at 10:56 AM
Cap Will Not Screw Into Place
Problem Cap not screwing into place. Cause Bent heater leads may be preventing the cap from centering. Solution Re-center the heater in the head a...
Tue, 13 May, 2014 at 4:52 PM
Software Issues
Problem Software issues Cause Various possible causes Solution See Eon™ Controller troubleshooting section for solutions to common software issues
Fri, 30 May, 2014 at 12:02 PM