Tempe™
RTD pins may be faulty
In order to determine if Tempe™ RTD pins are faulty, perform the following procedure:
Measure the resistance between pin A and ...
Thu, 29 May, 2014 at 4:17 PM
Heater pins may be faulty
In order to determine if Tempe™ heater pins are faulty, perform the following procedure:
Measure the resistance between pin ...
Fri, 30 May, 2014 at 10:18 AM
Problem
Crystals breaking prematurely Cause
Crystal not seated
properly Solution
Make sure that the crystal is seated properly inside the crystal co...
Fri, 30 May, 2014 at 11:04 AM
Problem
Crystal produces weak reading Cause
Crystal contact spring may have
become deformed Solution
If the contact spring has become
deformed, it ...
Fri, 30 May, 2014 at 11:11 AM
Problem
Heat conduction is insufficient to heat sensor head Cause
The heater’s ceramic
insulators may have become dislodged - which often occurs if the...
Fri, 30 May, 2014 at 10:56 AM
Problem
Cap not screwing into place. Cause
Bent heater leads may be preventing the
cap from centering.
Solution
Re-center the heater in the head a...
Tue, 13 May, 2014 at 4:52 PM
Problem
Software issues Cause
Various possible causes Solution
See Eon™ Controller troubleshooting section for solutions to common software issues
Fri, 30 May, 2014 at 12:02 PM